Self-Contained Steam SterilizerM9 / M9DM11 / M11DDescriptionPage 7ComponentOverviewPage 8Controls &IndicatorsPage 9Calling ForServicePage 33Operat

Seite 2 - MA616300i

8COMPONENTS OVERVIEWIllustration below shows location of sterilizer’s major components..DESCRIPTION OF COMPONENTS1. Reservoir Access Cover 5. Display2

Seite 3 - CONTENTS

9CONTROLS & INDICATORSThe following illustrations show the location of sterilizer’s controls and indica-tors. The chart describes their function.R

Seite 4 - Safety Instructions

103 POUCHES button a program cycle designed to process instru-ments in combination paper / plastic steriliza-tion pouches or wrapped instruments at 27

Seite 5

11OPERATIONSterilizer Preparation Before Operation1. Plug in sterilizer power cord. Follow instructions. Display shows:(*will display Model Number, M

Seite 6 - INSTALLATION

122. Open door and pour distilled water into fill opening (2) until the water level found in the Water Level Indica-tor Tube (3) is at the top of the

Seite 7

13Cleaning InstrumentsFollow item manufacturer’s and OSHA recommendations for handling and cleaning items.• As soon as possible after use, remove all

Seite 8 - Electrical Requirements

14• Place all containers so that the opening allows steam to enter, and air and condensate to drain from the container.• Sterilize jointed items in an

Seite 9 - OOR OPEN, CLOSE DOOR” is

15† Packs to have a minimum of 1/4 in. (6.4 mm) space between each other and away from all sterilizer surfaces.Operating the SterilizerPacks (†)4425 c

Seite 10 - COMPONENTS OVERVIEW

16Standard Cycle ParametersThe following table lists the standard cycle parameters for the various cycles.*Dry time can be changed from 0 to 60 minute

Seite 11 - CONTROLS & INDICATORS

17Standard Cycle OperationRefer to following steps for a detailed description of operating procedures:1. Display shows: 2. Close and latch door; ster

Seite 12 - Ref. Control Function

Owner’s Product Identification(information that you will need to provide for servicing - key information is highlighted)Model NumberSerial NumberName

Seite 13 - OPERATION

18 Press START. Sterilizer sounds a “beep” for two seconds to indicate cycle has started.• Filling Chamber: Display shows: Chamber automaticall

Seite 14 - EQUIPMENT

19• Automatic Door Open (Pertains only to M9/M11 UltraClave™)• Door To Open Display shows: An Audible signal is emitted to indicate to Operator door

Seite 15 - Cleaning Instruments

20Programming ModeThe following steps are for programming buttons 1 and/or 2 for applications that are not covered by standard cycle programs:Displays

Seite 16 - Maximum Capacities

21then displays: Press Displays: then Press changes VENT to FAST. Press changes VENT to SLOW.When in FAST vent, valve fully opens and

Seite 17

22Unloading Hot Trays and Cassettes (Using Tray/Cas-sette Tool)Tray Removal Insert tray (smaller) plate (1) into end of tool (2)Hook the top tab of t

Seite 18 - Standard Cycle Parameters

23Cassette RemovalInsert cassette plate (A) (larger plate) into end of the tool.Hook top, saw-toothed tab of tool to the top center of the cassette (3

Seite 19 - MA596700i

24List of Authorized AccessoriesListed below are the accessories which are authorized for use with these sterilizers. Unless noted, accessory can be u

Seite 20

252. Installation• Place back edge of tray in chamber. Pressing downward (A) on top of tray rack, slowly insert assembly into chamber (B).Daily1. Clea

Seite 21

26Weekly1. Clean Chamber and Trays • Drain water from the reservoir using drain tube located on front of unit.• Wash inside of chamber and trays with

Seite 22

27Perform Pressure Relief Valve Check(a) Press UNWRAPPED (b) Press START (c) Wait until pressure in chamber reaches 20 PSI (138 kPa).Display: Pu

Seite 23

CONTENTSIMPORTANT INFORMATION ... 2Intended Use o

Seite 24 - Tray Removal

28Non-Scheduled Maintenance1. Cleaning Fill / Vent and Air Filter Screens The filter screens (1) are intended to prevent debris from causing valve fai

Seite 25 - Cassette Removal

29Extended Use MaintenanceThe M9/M9D and M11/M11D are designed and tested to provide exceptional reliability throughout their service life. However, l

Seite 26 - Tray Rack & Plate

30Door gasket leaks door gasket is damaged or dirty.clean or replace door gasket (see Clean Door Gasket under Weekly maintenance).Items are not dry at

Seite 27 - EQUIPMENT ALERT

31Message GuideCAUTIONThe Message Guide list some of the messages that may occur during op-eration, along with possible causes and solutions.If an err

Seite 28

32C060: POWER UP MODESYSTEM HARDWAREPower was interupted briefly or an internal glitch.Unplug unit power cord for 1 minute and then plug back in. If

Seite 29 - XXX°F 20.0 PSI

33CALLING FOR SERVICEContact your Midmark Authorized Dealer, or log onto www.midmark.comModel and serial number information will be required when call

Seite 30 - Non-Scheduled Maintenance

34SPECIFICATIONSM9/M9D SterilizerPhysical Dimensions:Overall Length w/Plug ...20.38 in. (51.8 cm)Overall Width ...

Seite 31 - Troubleshooting Guide

35M11/M11D SterilizerPhysical Dimensions:Overall Length w/Plug ... 23.8 in. (60.5 cm)Overall Width...

Seite 32 - Maintenance

36PRINTER INSTALLATION AND OPERATIONThe printer is an optional add-on that some units may have. The printer may al-so be purchased as an accessory at

Seite 33 - Message Guide

376. Route loose end of PCB printer harness (1, Figure 2) through slot (A) in printer cav-ity (4) of top cover.7. Plug PCB printer har-ness (1) into p

Seite 34 - Message Guide - Continued

2IMPORTANT INFORMATIONIntended Use of ProductThe M9/M11 Ultraclave or M9D/M11D Autoclave can be used in medical and dental offices, hospitals, clinics

Seite 35 - CALLING FOR SERVICE

38Inserting the Paper Roll1. Remove the printer assem-bly (1,Fig. 4) from the printer cavity (A) and turn it onto its back.2. Remove paper spindle (2)

Seite 36 - SPECIFICATIONS

399. Turn the paper roll so as to take up the slack in the paper feeding to the printer. Make sure the roll of paper turns freely. If it does not tu

Seite 37 - M11/M11D Sterilizer

40Removing the Paper RollBefore removing the paper roll, advance the paper about 1 in. (2.5 cm) by press-ing down and holding the PAPER FEED button. L

Seite 38 - Installing the Printer

41When the sterilizer enters the Drying Phase the word “DRYING” is printed andthe printer will print the elapsed time in 5 minute increments starting

Seite 39 - Operating the Printer

42The figure below shows an example of a typical printout of a program cycle: Midmark M9 - v1.0.3 Total Cycles: 10

Seite 40 - Inserting the Paper Roll

43LIMITED WARRANTYSCOPE OF WARRANTYMidmark Corporation (“Midmark”) warrants to the original purchaser its new Alter-nate Care products and components

Seite 43

003-1249-00 Rev. Z (11/22/2011)Midmark Corporation60 Vista DriveP.O. Box 286Versailles, Ohio 45380-0286937-526-3662Fax 937-526-5542midmark.com To re

Seite 44

3ImportantInformationTransportation and Storage Conditions• Storage Temperature Range: ... -22°F to 140°F (-30°C to +60°C)• Relative Humidity...

Seite 45 - SF-1487 REV. A1

4INSTALLATIONLocation Requirements For SterilizerAdherence to the following recommendations for location of the sterilizer will contribute to optimum

Seite 46

5Support Surface Depth (C) - Support surface should be at least 21 in. (53.3 cm) (M11) or 18 in. (45.7 cm)(M9) deep. Allow at least 2 in. (5.1 cm) cl

Seite 47

6Re-Location Requirements For Sterilizer1. Disconnect power cord from electrical outlet and allow sterilizer to cool.2. Drain water from reservoir or

Seite 48

7DESCRIPTIONSafety FeaturesThe M9 / M11 UltraClave™ or M9D / M11D Autoclave sterilizers have the follow-ing Safety Features . . .• dry cycle door stop