10
Figure 1 – Diagnostics Procedure (M9 & M11 and M9D & M11D Style Key Pad Displays)
M9 & M11 Style Key Pad Display
M9D & M11D Style Key Pad Display
Step #3 continued from page 9.
Read the IMPORTANT note on page 8, before you
begin the following steps.
a) Press and hold the LIQUIDS and PACKS buttons
simultaneously while connecting the power cord into
an A.C. outlet.
b) Release the LIQUIDS
and PACKS buttons simultane-
ously to cause the display PC Board LED's to individ-
ually illuminate, and then extinguish one at a time in a
left to right sequence.
c) Press and release the buttons in the following order
and one at a time: UNWRAPPED, POUCHES, LIQ-
UIDS, PACKS, START, STOP and ON/STANDBY. When
each button is pressed, the corresponding LED will
illuminate and stay illuminated. When the sequence is
completed, the PRESSURE (PSI) display will display
the pressure that the control PC board is reading and
the TEMP (˚F)/TIME (MIN: SECONDS) display will dis-
play the temperature that the control PC board is read-
ing. When the displays register proper values, contin-
ue with calibration – proceed to Step #4, page 11.