Hardware and/or software problem – Watchdog timer reset error.
1. Poor electrical connection or faulty wiring (bad power source to
sterilizer).
2. Main PC board may be faulty.
Software interruption error.
• Run additional cycles. If problem continues, replace the Main PC
board.
Ram Test Error.
• Run additional cycles. If problem continues, replace the Main PC
board.
Door malfunction: Error Code 8 – Door switch did not return to normally
open condition after door pulse solenoid actuated. Error Code 9 –
Display Pressure exceeded 0.9 PSI during the Dry cycle.
• If pressure reading is not at zero (0), sterilizer needs to be re-calibrated.
Refer to 5.0 - Calibration Procedure.
• If door opens at end of cycle, check door switch continuity (normally
open).
• If door does not open at end of cycle, refer to 4.0 - Door Opening Test
Procedure.
• Inspect door mechanism for smooth operation. Lubricate if necessary.
• Check door switch and wire connections.
• Check door pulse solenoid.
Pressure display reads less than 24.5 psi.
• Check for pressure leaks. Refer to 1.0 - Leak Detection Guide.
• Temperature probe is dirty or malfunctioning.
• Sterilizer may need to be re-calibrated. Refer to 5.0 - Calibration
Procedure.
ERROR CODE 7 ERROR CODE 10
ERROR CODE 12
ERROR CODE 11
ERROR CODE 8 or 9
3.0 - ERROR CODES AND OTHER DIAGNOSTICS (continued)