Refer to the following steps for a detailed description of the program operating
procedures:
WARNING
Do not use this sterilizer in an explosive or oxygen-rich atmo-
sphere, or where flammable anesthetics are stored. To do so
could result in an explosion or fire.
Do not use this sterilizer for sterilizing volatile substances or for any
purpose other than its intended design. Burns and toxic or explosive
conditions could result.
Clean and dry instruments before putting them into the sterilizer.
Incomplete and improper cleaning of instruments will impede steriliza-
tion and will result in unsterile instruments which could lead to per-
sonal injury or death.
If the sterilizer malfunctions, immediately unplug sterilizer. If the steril-
izer continues to malfunction, call for service; do not attempt to repair
the sterilizer yourself.
Do not force door handle at any time. Chamber pressure may cause
door to open with extreme force. If the door handle does not move
freely, allow unit to cool and depressurize for 40 minutes before open-
ing door. Failure to adhere could result in serious personal injury or
death.
EQUIPMENT ALERT
Do not use toweling or packaging which may contain chlorine
bleach residue. Doing so could result in trays and/or chamber
rusting or discoloring. In extreme cases, the life of the chamber may be sig-
nificantly shortened.
NOTE
Use sterility monitors with each sterilization load. Also, if a sterilizing cycle is
terminated prematurely, reprocess the instruments to ensure the sterility of
the load.