18
• The sterilizer automatically begins processing its sterilization load.
• The FILLING lamp (5) illuminates and the chamber automatically fills to
the correct level.
• The STERILIZING lamp (6) illuminates, indicating that the sterilizer is in
the heating and sterilization phase of the selected cycle.
• When the correct temperature (3) and pressure (7) are reached, the ster-
ilizer begins its time countdown. The selected exposure time and temper-
ature are maintained throughout the sterilization phase.
• When the exposure time is complete, the sterilizer vents the chamber of
all steam and water back into the reservoir tank.
• When venting is complete and the pressure inside the chamber is
reduced to zero, the sterilizer signals the operator with five series of three
audible beeps, warning the operator that the sterilizer door is opening.
• The door opens, the DRYING lamp (8) illuminates, and the sterilizer
begins a 30 minute drying phase (unless Liquids cycle was selected).
• Leave the sterilization load in the chamber to dry or remove the steriliza-
tion load from the chamber if drying is not necessary.
• When the 30 minute drying phase is complete, the sterilizer signals the
operator that the drying phase is complete with five series of three audible
beeps and the COMPLETE lamp (9) illuminates to indicate that the pro-
gram cycle is complete.
• Remove the sterilizer load from the chamber.
5. Press the ON / STANDBY button (1) to return the sterilizer to the STANDBY
mode
or
if additional sterilization program cycles are desired, restart the
sterilizer at this time.
WARNING
Keep clear when the door is ready to open. Five series of
three audible beeps will be heard prior to partial door open-
ing. Leave door in this position until steam dissipates. Failure to do so
could result in severe burns from steam being released.
CAUTION
The processed load and metal inner surfaces will be hot at the
completion of the drying phase. Avoid contact. Failure to do so
could result in serious burns.
NOTE
The sterility of wrapped or packed instruments cannot be ensured unless the
wrapped or packed instruments are allowed to dry fully.