29
SPECIFICATIONS
Physical Dimensions:
Overall Length.................................559 mm (22 in)
Overall Width...................................457 mm (18 in)
Overall Height..................................445 mm (17.5 in)
Counter Area...................................457 mm x 406 mm (18 x 16 in)
Chamber .........................................279 mm Diameter x 457 mm depth
(11 in Diameter x 18 in depth)
Large Trays (2) ................................381 mm x 152 mm x 25 mm
(15 in x 9 in x 1 in)
Small Trays (2) ................................381 mm x 152 mm x 25 mm
(15 in x 6 in x 1 in)
Weight:
Empty Reservoir..............................45 Kgs (95 lbs)
With Shipping Carton......................48 kgs (105 lbs)
Water Reservoir Capacity
.............Approximately 4.54 Liters to Full Mark
(1.2 gallons)
Electrical Requirements
:
NOTE:
A separate (dedicated) circuit is recommended for this sterilizer. The
sterilizer should not be connected into an electrical circuit with other appliances
or equipment unless the circuit is rated for the additional load.
100 VAC, 50/60 Hz Model...............20 AMP Circuit, Single Phase
120 VAC, 60 Hz Model....................15 AMP Circuit, Single Phase
220-240 VAC, 50/60 Hz Model........10 AMP Circuit, Single Phase
Maximum Power Consumption
:
100 VAC, 50/60 Hz Model...............1425 Watts, 14.2 A @ 100V
120 VAC, 60 Hz Model....................1425 Watts, 11.9 A @ 120 V
220-240 VAC, 50/60 Hz Model........1500 Watts, 6.3 A @ 240 V
Chamber Pressure at
132°C (270°F)
.................................215 kPa (31 psi)
Safety Valve Setting
......................276 kPa (40 psi)
Certifications
.................................ISO-9001 Certified
LIMITED WARRANTY
SCOPE OF WARRANTY
Midmark Corporation (“Midmark”) warrants to the original purchaser its new Alter-
nate Care products and components (except for components not warranted under
“Exclusions”) manufactured by Midmark to be free from defects in material and
workmanship under normal use and service. Midmark’s obligation under this war-
ranty is limited to the repair or replacement, at Midmark’s option, of the parts or the
Operator
Maintenance
Calling For
Service
Specifications
Limited
Warranty
Return To Table Of Contents