23
3. Cleaning Chamber Filter:
(a) Before performing this pro-
cedure, make sure that the
sterilizer has cooled to room
temperature.
(b) Open the sterilizer door and
remove all trays, tray rack,
and tray plate from the
chamber. Locate the cham-
ber filter (1) on the bottom of
the chamber.
(c) Grasp the filter (1) and gently pull upwards while twisting slightly (a pair
of pliers may be used if filter is stuck). The filter may be cleaned with
mild soap and distilled or demineralized water. A small stiff bristled
brush or ultrasonic cleaner may be helpful to remove foreign objects
from the filter surface. Rinse the filter with distilled or demineralized
water.
(d) Replace the filter (1) by inserting the filter into the hole in the chamber
bottom and pressing downwards while twisting slightly.
Quarterly
1. Remove and Clean Door Gasket - the door gasket and dam gasket must be
removed and checked and the gasket housing channel must be cleaned as
follows:
NOTE
If the filter cannot be cleaned ef-
fectively, the filter must be re-
placed.
EQUIPMENT ALERT
Do not operate the sterilizer without the filter in place. Doing so
could result in the sterilizer malfunctioning.
1
Operator
Maintenance
Return To Table Of Contents