program cycle is complete by sounding five audible “beeps”. The sterilizer
may be switched to standby at this time or restarted if additional sterilization
programs are desired.
Mechanical Features
The M11UltraClave™ sterilizers . . .
• have a tray rack which can hold two small and two large trays. The tray rack
can be removed for periodic cleaning of the chamber, tray plate, tray rack,
and filter.
• have an optional printer which provides a permanent record of actual expo-
sure times and temperatures.
• have a removable power supply cord (100 and 230 VAC units only).
• have a two gaskets in the door; a dam gasket and a door gasket which can
easily be replaced by operator in seconds.
Safety Features
The M11UltraClave™ sterilizers . . .
• have a
door stop
which prevents the door from opening fully if there is a
slight residual pressure in the chamber when the door handle is operated.
The door stop also retains the door in a partially open position during the
drying cycle.
• have a
door closed switch
which is used to signal the sterilizer that the
door is completely closed. This switch is activated just as the door becomes
fully closed. An audible signal will sound and the “Door Ajar” light will flash if
the door is not fully closed.
•have a
water low indicator
which signals the operator if there is insufficient
water to fill the chamber to the proper level. An audible signal will sound and
the “Water Low” light will flash if the reservoir needs to be refilled with water.
• have their
chamber temperature monitored
during a cycle to prevent an
overheat condition. If chamber temperature reaches or exceeds 146°C
(295°F), power to the sterilizer is cut off.
• have a
pressure relief valve
which opens to provide backup protection to
reduce chamber pressure in the event that the chamber steam pressure
exceeds the allowable limit. Released steam is directed out of the bottom of
the unit’s cabinet. An access cover is located on the top cover for relief valve
operational checks.
• have a
fault detection circuit
which monitors all functions of the sterilizer
during a cycle. If a fault occurs during a cycle, the sterilizer monitoring cir-
cuitry will stop the cycle, sound an audible signal, and light a flashing error
lamp.